Ambix activ Pump Sets
The activ sets are specially designed for the Ambix activ. Every set includes a sophisticated particle/air filter which, in conjunction with the air alarm sensor of the Ambix activ, prevents air administration. Downstream kink protection is also integrated to reduce the risk of kink especially in ambulatory use. Free flow protection is provided by the activ clamp, an advanced second generation pinch clamp. This clamp automatically closes when the pump door is opened, thereby preventing free-flow. An integrated flow cap facilitates priming proceedings.
Note: All activ sets are for single use only. Note: Fresenius Kabi usually recommends placing the container 50 cm (± 10 cm) above the pump. Note: For further information about activ sets, please refer to the related instructions for use.
Warning! In case of use of ASV, the pressure level may be changed (see page 4).
Warning! If the reservoir is placed above the injection site, ensure that the clamp is closed or disconnected from patient.
The Ambix activ is designed to operate exclusively with the appropriate Fresenius Kabi activ sets including the patented activ clamp (see ordering information on page 43). The use of other sets is prohibited and may endanger patient safety either through the risk of air infusion, reduced flow rate accuracy or free-flow conditions. |
|